The RSI aerosol generators produce an industry standard true Laskin III-A repeatable aerosol and feature adjustable aerosol concentration output to meet multiple application requirements.
These generators are used in conjunction with an aerosol photometer to measure the aerosol penetration of an individual filter or filter system, typically to identify, quantify and repair leaks. Aerosol concentration output varies by generator model however concentrations can be achieved from 100ug/l down to 10ug/l at airflows ranging from 400405 to 8,100 cfm*.
Model | RSI 6N | RSI 3N |
---|---|---|
Size | 11" L x 11" W x 10" H | 11" L x 8" W x 10" H |
Weight | 7.5 Kg | 5.5 Kg |
Aerosol Output | 5060-8100 CFM | 50–4000 CFM |
Concentration | 100 UG/L @ 810CFM100 UG/L @ 4000 CFM | 10 UG/L @ 8100 CFM10 UG/L @4000 CFM |
Generator Type | 1 to 6 Nozzle | 1 to 3 Nozzles |
Compressed Air | Min 2.65 CFM/ nozzle @ 20 PSI | |
Aerosol Type | Cold Type Poly-disperesed |